The Portable Photoionization Detector PPID3 is a detector for volatile organic compounds (VOCs) with an ionisation potential < 10.6 eV. The use of new technologies for the light source and the sensor enables a high stability of the measurement and longer maintenance intervals. |
Product Specification
Extended measuring rangeNewA new sensor unit allows three staggered measuring ranges up to 2 000 ppm (isobutene) with high resolution at low concentrations. |
Easy user interfaceA graphic display with intuitive user guidance, clear text instructions and step-by-step configuration support the ease of use. |
High performance PID lampThe ionisation source is a hollow cathode lamp with an energy of 10.6 eV. The Ceramic Discharge Channel ensures high stability and a longer lifetime of up to 15 000 hours and more. |
Acoustic and visual warning functionsThe Portable PID has an acoustic alarm and two ultrabright LEDs that warn when adjustable concentration values are exceeded. |
Simple calibrationOnly one calibration gas is required for the simple menu driven two-point calibration. |
Diverse charging optionsThe Portable PID is equipped with a standard USB-C connection and can be charged with any common USB-C wall adapter or car charger and has the option for wireless charging. |
Temperature / humidity compensationThe integrated humidity / temperature sensor measures the physical environment data in addition to the sensor signal and compensates their influence on the measuring result. |
Technical Data
| Detector principle |
VUV-Photoionization with 10.6 eV hollow cathode lamp |
| Detection ranges | 0 … 2000 ppm Isobutene * |
| Display range | 0 … 20 000 ppm, depending on response factor of detected substance |
| Lower detection limit | Typical 0.05 ppm * |
| Display resolution | Dynamic |
| Response time | T90 < 10 s * |
| Signal integrity | up to 100 ppm typical > 98 % * up to 2 000 ppm typical > 95 % * |
| Influence of humidity and temperature |
Humidity and temperature compensation at 0 ... 50 °C and 0 … 90 % rH |
| Operating conditions | -10 … 50 °C 0 … 95 % rH, non-condensing |
| Storage conditions | -20 … 60 °C 0 … 95 % rH, non-condensing (45°C for 3 month limited by battery) |
| Gas sampling | Integrated diaphragm pump (about 200-300 ml/min) with function monitor, Sample inlet with dust and water protection filter |
| PID lamp lifetime | 10.6 eV: Min. 8 000 h, typical more than 15 000 h |
| Alarms | Locked alarm, self-triggering alarm, 2 adjustable alarm levels |
| Power supply | Integrated lithium-ion battery 3,6 V/24 Wh |
| Signalisation | Visual: Ultra-bright LEDs, Acoustic: Alarm, at least 90 dB (A) in 30 cm |
| Runtime Charging time |
Runtime >15 hours Charge approx. 6 hours with USB-C charger (5V / 15W), approx. 10 hours with Wireless-Charger |
| Display | Full graphic monochrome OLED-display |
| Digital interface | USB-C |
| Calibration | Two-point calibration zero gas and span gas via sample inlet |
| Response factors |
about 100 selectable response factors can be set |
| Memory | For 8 hours one measurement data per second including time, temperature, and humidity |
| Program functions | Continuous measurement with automatic and manual data storage, threshold monitoring, two-point calibration, self-test, fresh air compensation |
| Dimension, weight | 220 mm x 75 mm x 48 mm (L x W x H), less 450 g |
| Ingress Protection | IP54 |
| Warranty | 2 years, except for components in gas path and wearing parts |
| Approvals | EMC directive 2014/30/EU |
| * The indicated values were obtained under standardized conditions with 10.6 eV lamp. Test gas was isobutene in synthetic air. |